Refractive Surgerycomplex Case Management | Aug 2016

Hyperopic Shift With Corneal Inlay

Case Presentation

A 52-year-old woman presented with symptomatic presbyopia and a desire for near vision correction. After a discussion of the risks and benefits of a corneal inlay as well as alternatives, the patient chose to receive the Kamra (AcuFocus). She has a history of fibromyalgia and was recently diagnosed with celiac gluten sensitivity.

Preoperatively, the patient had a distance UCVA of 20/15, a manifest refraction of plano, a cycloplegic refraction of +0.25 D, and a near UCVA of 20/400 in her right eye. In her dominant left eye, she had a distance UCVA of 20/15, a manifest refraction of plano, a cycloplegic refraction of plano, and a near UCVA of 20/400. Corneal pachymetry readings were 582 µm OD and 587 µm OS.

A presurgical evaluation with the AcuTarget HD (Visiometrics) was unremarkable, with an angle kappa of 91 µm, a mean ocular scatter index (OSI) of 0.38, and Optical Quality Analysis System accommodative range of 0.50 D.

The surgeon used the iFS Laser (Abbott) to create a 260-µm corneal pocket (44% of corneal depth) with a spot separation and line separation of 4 µm in the patient’s nondominant right eye. The plan was to place the corneal inlay on the Purkinje axis (within 300 µm). A postoperative evaluation with the AcuTarget HD showed that the implant was 115 µm nasal to the Purkinje light reflex.

One week after surgery, the patient began complaining of extreme photophobia in her right eye and poor near visual acuity (20/100). An examination at that time was unremarkable except for mild corneal edema. The doctor prescribed prednisolone acetate (Pred Forte; Allergan) six times per day. At the 2-week postoperative visit, the patient reported no improvement in her symptoms as well as continued photosensitivity and poor near vision. In addition, the IOP had risen to 25 mm Hg, and her distance manifest refraction was +0.75 = 20/25. The surgeon prescribed a fixed combination of brimonidine tartrate and timolol maleate (Combigan; Allergan), and the steroid was continued at six times per day.

Figure 1. The AcuTarget HD preoperative planning view demonstrates a Purkinje image (green X) versus the center of the pupil (yellow X) and thus the amount of angle kappa. In this case, angle kappa cord length measures 97 μm (A). Four months after surgery, the AcuTarget HD assessment shows the center of the Kamra inlay (red X) in comparison to the Purkinje image (green X). Ideally, the difference between these points is less than 300 μm. In this case, it measures 115 μm nasal and 8 μm superior (B).

Figure 2. The preoperative OSI measured by the double-pass retina image analysis of the AcuTarget HD was 0.2 (A). Four months postoperatively, the OSI had increased to 1.3 (B).

Six weeks after surgery, the patient reported a slight improvement in vision and decreased photosensitivity. Near UCVA measured 20/60. The steroid was tapered over 3 months. At the 3-month postoperative visit, the distance UCVA was 20/30, the manifest refraction was +1.00 D = 20/20, the near UCVA was 20/70, and the IOP measured 22 mm Hg. The patient started cyclosporine ophthalmic emulsion 0.05% (Restasis; Allergan) to treat potential occult dry eye. Four months postoperatively (1 month after steroid cessation), the manifest refraction was +1.50 D, correctable to 20/20, and near UCVA had dropped to 20/80. The examination was unremarkable. The patient continued the cyclosporine, but steroids were not restarted for fear of a steroid-induced pressure response.

Five months after surgery, the patient presents with a distance UCVA of 20/70 and a near UCVA of 20/100. The manifest refraction is +2.25 D sphere = 20/20 OD. An examination is unremarkable, with no signs of corneal inflammation or thinning. An assessment with the AcuTarget HD shows a well-centered corneal inlay. The OSI measures 1.3, and the Optical Quality Analysis System accommodative range is 1.00 D. Corneal pachymetry by tomography measures 602 µm (Figures 1-4).

What is the etiology of the refractive shift, and what are your thoughts on preventative measures based on the patient’s preoperative presentation? What would your next step be in terms of medical and/or surgical therapy? If medical therapy were not effective, how would you proceed?

—Case prepared by William F. Wiley, MD.

Figure 3. Preoperative topography with the Pentacam showed normal corneal shape and thickness.

Figure 4. Four months postoperatively, the topographic shape is normal, and there is no apparent central elevation or “red ring” that can be indicative of aggressive corneal wound healing with an accompanying hyperopic shift.

LANCE KUGLER, MD, PCEO

The hyperopic shift of +2.25 D is consistent with an aggressive wound healing response, as is the characteristic “red ring” seen on Placido disc topography, suggestive of peripheral edema. Other symptoms of aggressive wound healing such as photophobia are not present. Nor are slit-lamp examination findings such as corneal haze or edema. The OSI increased from 0.38 preoperatively to 1.3 at 5 months, and pachymetry increased from 582 to 602 µm, suggestive of mild corneal edema that may not be evident on examination. The presence of hyperopia greater than 1.00 D at 3 months suggests that the wound healing response was active at that time and that perhaps restarting steroids would have slowed or reversed the process.

Now at 5 months (2 months after steroid cessation) with a progressive hyperopic shift, mild corneal edema, and no evidence of haze or an infiltrate, it would be reasonable to resume the postoperative steroid regimen while continuing the fixed combination of brimonidine and timolol and to treat the ocular surface aggressively. If under close observation the hyperopia, edema, or topography worsened or did not improve, then I would explant the inlay. If the aggressive wound healing response resolves but the hyperopia persists, then explanation is still warranted, because any further enhancement of refractive error such as hyperopic PRK is best avoided in an eye with this history.

R. LUKE REBENITSCH, MD

This is an interesting case of a usually treatable complication during the postoperative period following implantation of the Kamra. Hyperopic shifts are thought to be the result of keratocyte activation behind the inlay, resulting in a “red ring” on Placido disc topography. They typically occur at least a month after the procedure during the steroid taper, however, rather than within the first few weeks. This patient had a much more robust response to the device, which caused early photophobia and a hyperopic shift.

This patient would have appeared to be an excellent candidate for the Kamra. She had an OSI that was very low at 0.2, exhibited no significant ocular dryness, had lost most accommodative amplitude, and possessed a cornea thick enough for implantation. My initial strategy for management would have been the same as described in the case presentation: increase topical steroids and use serial topographies and midpoint refractions to monitor the response. Unfortunately, this patient was a steroid responder, complicating postoperative management. Given the progressive hyperopic shift and increased IOP, I would recommend explantation of the device and would expect no significant long-term sequelae.

Preventative measures for this ideal candidate are limited. New research, however, suggests that deeper implantation of the Kamra may mitigate some of the risk of a hyperopic shift from the decreased density of keratocytes in the posterior cornea. For a patient with a cornea as thick this patient’s measured with the Pentacam (Oculus Surgical), a pocket of approximately 50% depth may prevent some—or possibly all—of the hyperopic shift she experienced.

GEORGE O. WARING IV, MD

This patient appears to be a great candidate for a Kamra inlay given the refraction and excellent retinal image quality, both statically and dynamically, as determined by double-pass wavefront imaging. These findings are consistent with a diagnosis of stage 1 dysfunctional lens syndrome and a healthy ocular surface.

Her past medical history may give some clues to a nonspecific pro-inflammatory state that we might appreciate retrospectively. Small-aperture inlays tend to perform best with a -0.75 D spherical target, which optimizes near performance without sacrificing distance vision and “cushions” near vision in the event of refractive fluctuation such as is seen here.

Tight spot and line femtosecond laser settings and an adequate low-dose steroid taper over 3 months help to minimize refractive fluctuations; both were used in this case. Careful attention should be paid to IOP for a steroid response, as was observed in this case, resulting in early cessation of the steroid. In most cases, appropriate dosage and duration of a mild steroid will normalize a refractive shift over time.

If a hyperopic shift worsens despite appropriate steroid dosing and duration, thus affecting the patient’s near and distance UCVA and satisfaction, or if the IOP is intractably high for the optic nerve status due to a steroid response, then the inlay should be removed. The removability of the Kamra in the rare event that it is indicated is a true benefit of this “additive” technology. In the US FDA investigational device exemption trial, all eyes that underwent inlay removal retained BCVA.

Section Editor Alan N. Carlson, MD
• professor of ophthalmology and chief, corneal and refractive surgery, Duke University Eye Center, Durham, North Carolina

Section Editor Stephen Coleman, MD
• director of Coleman Vision, Albuquerque, New Mexico

Section Editor Karl G. Stonecipher, MD
• clinical associate professor of ophthalmology, University of North Carolina, Chapel Hill
• director of refractive surgery, TLC in Greensboro, North Carolina

Section Editor William F. Wiley, MD
• private practice at Cleveland Eye Clinic, Cleveland, Ohio
• (440) 526-1974

Lance Kugler, MD, PCEO
• surgeon and CEO of Kugler Vision, Omaha, Nebraska
• director of refractive surgery, University of Nebraska Medical Center
lkugler@kuglervision.com
• financial interest: none acknowledged

R. Luke Rebenitsch, MD
• refractive surgeon, ClearSight Center, Oklahoma City, Oklahoma
• (405) 733-2020; dr.luke@clearsight.com
• financial disclosure: has received speaking and educational honoraria from AcuFocus

George O. Waring IV, MD
• director of refractive surgery and associate professor of ophthalmology, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina
n medical director, Magill Vision Center, Mt. Pleasant, South Carolina
georgewaring@me.com; Twitter @georgewaring
• financial disclosure: consultant to AcuFocus and Visiometrics

Advertisement - Issue Continues Below
Publication Ad Publication Ad
End of Advertisement - Issue Continues Below

NEXT IN THIS ISSUE