We noticed you’re blocking ads

Thanks for visiting CRSToday. Our advertisers are important supporters of this site, and content cannot be accessed if ad-blocking software is activated.

In order to avoid adverse performance issues with this site, please white list https://crstoday.com in your ad blocker then refresh this page.

Need help? Click here for instructions.

Refractive Surgery: Feature Story | Sep 2010

Wavefront-Optimized Procedures Associated With Less Postoperative Spherical Aberration Versus Wavefront-Guided

My hypothesis regarding the effectiveness of the wavefront-optimized algorithm was reflected in my results.

The approval of customized laser refractive surgery in the United States in 2002 (Alcon Laboratories, Inc., Fort Worth, TX) and 2003 (Abbott Medical Optics Inc., Santa Ana, CA, and Bausch + Lomb, Rochester, NY) was accompanied by promises of “super” vision. Although the results with these systems were as good as or better than those obtained to date with conventional laser surgery, the promise of super vision has yet to be realized.

I believe that the biggest reason for this shortfall, if sphere and cylinder have been corrected, is that it is much easier to measure optical aberrations than it is to accurately correct them. The presence of clinically significant optical aberrations in the majority of eyes is generally minimal, with the exception of spherical aberration. Small variations in the pupil’s size and position, and variable alignment with the aberrometer, will introduce measurable changes to the wavefront measurement, all of which tend to diminish the surgery’s effectiveness in reducing aberrations.

With these considerations in mind, I began using the WaveLight Allegretto (Alcon Laboratories, Inc.) 400-Hz excimer laser platform. The Gaussian beam profile, eyetracking system, and speed of ablation (less than 1 second per diopter on average for a myopic correction) were designed to minimize intraoperative variability in the refractive correction. I believed that, with these advantages and the wavefront-optimized algorithm to eliminate the induction of spherical aberration, I was likely to see low induction of postoperative higher-order aberrations (HOAs). I am happy to say that my results to date have met my expectations. I decided to test my theory that aberrations were as low or lower with the WaveLight system using the wavefront-optimized algorithm as they were with the Visx Star S4 using the CustomVue wavefront-guided algorithm (Abbott Medical Optics Inc.) by comparing a sample of eyes treated on both platforms. My center has both laser platforms.

DATA COMPARISON
Because the aberrometer’s measurements can vary, I used a single instrument (WaveScan; Alcon Laboratories, Inc.) to measure pre- and postoperative aberrations on a series of 42 eyes. Of these eyes, 19 were treated with the Visx Star S4 with iris registration using the CustomVue wavefront-guided algorithm. The remaining 23 eyes were treated using the Allegretto wavefront-optimized algorithm. All eyes were myopic, except for two mixed astigmatism cases that were treated with the Allegretto system. Mean sphere was approximately -3.00 D for both groups, with a maximum treatment of -6.50 D. Average cylinder treatment was -0.75 D for both groups, with a maximum treatment of about -2.50 D. There was no statistically significant difference between the refractive errors in the two groups preoperatively, as determined by a t-test. Because the wavefront diameter affects the wavefront measurement, I also tested to see if there was any statistical difference between the diameter of the readings between the groups; there was not. Total HOAs were slightly higher in the CustomVue patients, but spherical aberration was not statistically significantly different between groups.

All eyes were refracted 1-month postoperatively. The postoperative sphere and cylinder corrections were not statistically different between the two groups of patients; the average sphere was 0.30 D in the Allegrettotreated eyes and 0.40 D in the Visx-treated eyes. The average cylinder was -0.40 D for eyes treated with both systems, and the average spherical equivalent was 0.09 for the Allegretto-treated and 0.21 for the Visx-treated eyes. All eyes were also measured again on the WaveScan system 1 month after surgery. I compared the postoperative readings with the preoperative readings for each system using a t-test.

The pre- to postoperative change in total HOAs for all eyes is shown in Figure 1. The postoperative measurements obtained with the Allegretto wavefront-optimized procedure for total HOAs and spherical aberration were statistically equivalent (total HOAs, P = .82; spherical aberration, P = .30). The results using the CustomVue wavefront-guided technique were not, however. Specifically, total HOAs with the CustomVue treatment were higher after surgery (P = .015), with most of that difference attributable to a large increase in spherical aberration (preoperative, 0.11 µm; postoperative, 0.34 µm; P = .002).

INTERPRETATION OF RESULTS
It was reassuring to see that my hypothesis regarding the effectiveness of the wavefront-optimized algorithm was borne out by the data I collected. One could argue that a system that measures and treats based on detailed HOAs has the potential to perform a superior ablation, but the delivery system significantly influences results. The Allegretto wavefront-optimized algorithm reduces the induction of spherical aberration during surgery, a function of both the algorithm and the high-speed, small-spot laser used to perform the ablation. This is reflected in the negligible increase in HOAs observed in my patient sample. The CustomVue platform performed similarly with regard to lower-order aberrations (sphere and cylinder) but did not demonstrate the ability to treat the HOAs measured preoperatively. Spherical aberration is the primary contributor to troublesome total HOAs. It appears, in this data set, that the ability to effectively manage spherical aberration at the time of surgery (ie, with the Allegretto platform) is more important than the measurement and attempted correction of more detailed HOAs (ie, with the CustomVue platform).

THE BOTTOM LINE
The most important feature of any excimer laser system is its ability to accurately and predictably treat a wide range of refractive errors. Most current laser systems can meet that goal. The ability to control optical aberrations is the next important criterion. In my experience, the Allegretto wavefront-optimized algorithm has demonstrated an ability to eliminate the induction of spherical aberration, thereby reducing or eliminating any increase in HOAs. Other authors have found no difference between the HOAs from the Visx Star S4 and the WaveLight-optimized treatment1 and wavefront-optimized versus other wavefront-guided treatment (eg, CustomCornea).2 Still others have noted that HOAs were lower postoperatively as a percentage reduction from preoperatively with the Visx (although there was a more restrictive limit on preoperative HOAs treated with the WaveLight platform).2

It would be a useful exercise in your practice to determine if the wavefront-guided treatment algorithm of your laser system is producing a similar effect. The ability to measure and then treat the wavefront is crucial to success. If your wavefront-guided system is unable to do either or both in a predictable manner, then your results may not be as “custom” as you would like.

Fred Mattioli, MD, practices at the LasikPlus center in Houston. He acknowledged no financial interest in the products or companies mentioned herein. Dr. Mattioli may be reached at 713-355-3500; fmattioli@lca.com.

  1. Perez-Straziota CE,Randleman JB,Stulting RD.Visual acuity and higher-order aberrations with wavefront-guided and wavefront-optimized laser in situ keratomileusis.J Cataract Refract Surg.2010;36(3):437-441.
  2. Durrie DS,Smith RT,Waring GO 4th,et al.Comparing conventional and wavefront-optimized LASIK for the treatment of hyperopia.J Refract Surg.2010;26(5):356-363.
  3. Binder PS,Rosenshein J.Retrospective comparison of 3 laser platforms to correct myopic spheres and spherocylinders using conventional and wavefront-guided treatments.J Cataract Refract Surg.2007;33(7):1158-1176.
Advertisement - Issue Continues Below
Publication Ad Publication Ad
End of Advertisement - Issue Continues Below

NEXT IN THIS ISSUE